INCREASING OF ACCURACY FOR DENSITY MEASUREMENT MAKING BY A RADIOISOTOPIC BASIS WEIGHT SENSOR

The experimental investigations of spatial energy distribution for beta particles in a measuring gap of a basis weight sensor are described. We propose the measurement procedure, which employs a local domain of the ionization chamber input window. The average scattering angles in polyethylenterephta...

Full description

Saved in:
Bibliographic Details
Main Authors: O. L. Ermakovich, G. A. Lisovsky, P. V. Kuchynski, I. A. Titovitsky
Format: Article
Language:English
Published: Belarusian National Technical University 2015-08-01
Series:Приборы и методы измерений
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/207
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The experimental investigations of spatial energy distribution for beta particles in a measuring gap of a basis weight sensor are described. We propose the measurement procedure, which employs a local domain of the ionization chamber input window. The average scattering angles in polyethylenterephtalate films are determined. Suggested technical solutions allow more then three times to reduce the error of measuring of radioisotope densitometer.
ISSN:2220-9506
2414-0473