Acoustic System of Determining the Instantaneous Volume of the Blood Part of the Ventricular Assist Device POLVAD-EXT
The paper presents the results of investigations concerning the noninvasive method of estimating the actual volume of the blood chamber of the POLVAD-EXT type ventricular assist device (VAD) during its operation. The proposed method is based on the principle of Helmholtz’s acoustic resonance. Both...