On acceptable values of disregard limits for impurities and signal-to-noise ratio when checking chromatographic system sensitivity

The article provides answers to urgent questions relating to the acceptable values of disregard limits for impurities and signal-to-noise ratio ( S / N ) when checking the sensitivity of the chromatographic system. It gives recommendations on the «typical» and maximum acceptable values of disregard...

Full description

Saved in:
Bibliographic Details
Main Author: N. A. Epshtein
Format: Article
Language:Russian
Published: Federal State Budgetary Institution ‘Scientific Centre for Expert Evaluation of Medicinal Products’ of the Ministry of Health of the Russian Federation (FSBI ‘SCEEMP’) 2018-02-01
Series:Регуляторные исследования и экспертиза лекарственных средств
Subjects:
Online Access:https://www.vedomostincesmp.ru/jour/article/view/124
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The article provides answers to urgent questions relating to the acceptable values of disregard limits for impurities and signal-to-noise ratio ( S / N ) when checking the sensitivity of the chromatographic system. It gives recommendations on the «typical» and maximum acceptable values of disregard limits for impurities. The article also demonstrates that the requirement of S / N ≥ 10 can be unnecessarily rigid in some cases: there could be found convincing arguments justifying a smaller S / N value for the sensitivity solution, for example, in cases when an identified impurity is determined by a specific procedure and the impurity peak appears before the peak of the main substance.