Development and validation of a HPLC method for the determination of metronidazole, oxytetracycline and furazolidone in veterinary formulations

An isocratic reversed phase high performance liquid chromatographic method with DAD detection was developed for the analysis of metronidazole, oxytetracycline and furazolidone. The mobile phase consisted of pH 2.5 phosphate buffer solution, methanol and acetonitrile (70:18:12). The UV detection was...

Full description

Saved in:
Bibliographic Details
Main Authors: Violeta Tauber (Giugiu), Viorica Chiurciu
Format: Article
Language:English
Published: Asociaţia Naţională a Fabricanţilor de Produse de Uz Veterinar 2014-06-01
Series:Medicamentul Veterinar
Online Access:http://www.veterinarypharmacon.com/docs/1345-2014-VD-8-1-ART.9.EN.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An isocratic reversed phase high performance liquid chromatographic method with DAD detection was developed for the analysis of metronidazole, oxytetracycline and furazolidone. The mobile phase consisted of pH 2.5 phosphate buffer solution, methanol and acetonitrile (70:18:12). The UV detection was carried out at 264 nm, and the flow rate was 1.100 mL/min. The separation was carried out on a Nucleosil C18 column, 5 µm 250 mmx4.6 mm, which was maintained at 25 0C. This method was validated by system suitability parameters, linearity, limits of detection and quantification, precision and accuracy
ISSN:1843-9527
2069-2463