Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices

A digital micromirror device (DMD) micromirrors periodic spatial structure is a measuring scale in interior orientation parameters calibration of optoelectronic devices problems, when using a DMD as a testobject. It is important that DMD micromirrors periodic spatial structure remains constant. Chan...

Full description

Saved in:
Bibliographic Details
Main Author: M. A. Starasotnikau
Format: Article
Language:English
Published: Belarusian National Technical University 2020-06-01
Series:Приборы и методы измерений
Subjects:
Online Access:https://pimi.bntu.by/jour/article/view/649
Tags: Add Tag
No Tags, Be the first to tag this record!
_version_ 1839572186460848128
author M. A. Starasotnikau
author_facet M. A. Starasotnikau
author_sort M. A. Starasotnikau
collection DOAJ
description A digital micromirror device (DMD) micromirrors periodic spatial structure is a measuring scale in interior orientation parameters calibration of optoelectronic devices problems, when using a DMD as a testobject. It is important that DMD micromirrors periodic spatial structure remains constant. Change in a DMD micromirrors spatial structure may occur due to heating. In addition to heating a DMD, an optoelectronic device photodetector is also subject to heating and, accordingly, change in its spatial structure. It is necessary to estimate change in a spatial structure of DMD micromirrors and an optoelectronic device photodetector.A DMD micromirrors spatial drift and a DMD micromirrors spatial drift together with a digital camera photodetector pixels spatial drift for operation 4 h are analyzed. The drift analysis consisted in the points array position assessing formed by a DMD and projected onto a digital camera. When analyzing only a DMD micromirrors drift, a digital camera was turned on only for shooting time for exclude digital camera influence. A digital camera did not have time to significantly heat up, during this time. After a digital camera it cooled to a room temperature.Average drift of all DMD micromirrors determines the accuracy of interior orientation parameters calibration of optoelectronic devices using a DMD in time. Maximum drift of all micromirrors after switching on is observed. Minimum DMD warm-up time is 60 min for average drift of all micromirrors less than 1 μm is necessary. Minimum DMD warm-up time is 120 min when using a DMD together with a digital camera is necessary.A DMD expansion uniformity determines the accuracy of interior orientation parameters calibration of optoelectronic devices using a DMD, because irregular expansion disturbs micromirrors periodicity. The average change in distance of neighboring points is less than 0.1 μm for every 20 min.Thus, a DMD can be used as a test-object in interior orientation parameters calibration of optoelectronic devices. The results can be used as compensation coefficients of change in DMD micromirrors spatial structure due to temperature effects during operation, if more accurate are necessary.
format Article
id doaj-art-a2b6b6b6c4ec4a2a9b09204a6b5196f0
institution Matheson Library
issn 2220-9506
2414-0473
language English
publishDate 2020-06-01
publisher Belarusian National Technical University
record_format Article
series Приборы и методы измерений
spelling doaj-art-a2b6b6b6c4ec4a2a9b09204a6b5196f02025-08-04T14:24:20ZengBelarusian National Technical UniversityПриборы и методы измерений2220-95062414-04732020-06-0111212213110.21122/2220-9506-2020-11-2-122-131510Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic DevicesM. A. Starasotnikau0Peleng JSCA digital micromirror device (DMD) micromirrors periodic spatial structure is a measuring scale in interior orientation parameters calibration of optoelectronic devices problems, when using a DMD as a testobject. It is important that DMD micromirrors periodic spatial structure remains constant. Change in a DMD micromirrors spatial structure may occur due to heating. In addition to heating a DMD, an optoelectronic device photodetector is also subject to heating and, accordingly, change in its spatial structure. It is necessary to estimate change in a spatial structure of DMD micromirrors and an optoelectronic device photodetector.A DMD micromirrors spatial drift and a DMD micromirrors spatial drift together with a digital camera photodetector pixels spatial drift for operation 4 h are analyzed. The drift analysis consisted in the points array position assessing formed by a DMD and projected onto a digital camera. When analyzing only a DMD micromirrors drift, a digital camera was turned on only for shooting time for exclude digital camera influence. A digital camera did not have time to significantly heat up, during this time. After a digital camera it cooled to a room temperature.Average drift of all DMD micromirrors determines the accuracy of interior orientation parameters calibration of optoelectronic devices using a DMD in time. Maximum drift of all micromirrors after switching on is observed. Minimum DMD warm-up time is 60 min for average drift of all micromirrors less than 1 μm is necessary. Minimum DMD warm-up time is 120 min when using a DMD together with a digital camera is necessary.A DMD expansion uniformity determines the accuracy of interior orientation parameters calibration of optoelectronic devices using a DMD, because irregular expansion disturbs micromirrors periodicity. The average change in distance of neighboring points is less than 0.1 μm for every 20 min.Thus, a DMD can be used as a test-object in interior orientation parameters calibration of optoelectronic devices. The results can be used as compensation coefficients of change in DMD micromirrors spatial structure due to temperature effects during operation, if more accurate are necessary.https://pimi.bntu.by/jour/article/view/649dmdoptoelectronic deviceinterior orientation parametersphotodetectortemperature drift
spellingShingle M. A. Starasotnikau
Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices
Приборы и методы измерений
dmd
optoelectronic device
interior orientation parameters
photodetector
temperature drift
title Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices
title_full Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices
title_fullStr Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices
title_full_unstemmed Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices
title_short Assessment of Temperature Effects in Interior Orientation Parameters Calibration of Optoelectronic Devices
title_sort assessment of temperature effects in interior orientation parameters calibration of optoelectronic devices
topic dmd
optoelectronic device
interior orientation parameters
photodetector
temperature drift
url https://pimi.bntu.by/jour/article/view/649
work_keys_str_mv AT mastarasotnikau assessmentoftemperatureeffectsininteriororientationparameterscalibrationofoptoelectronicdevices