Bench studies of a small axial pump for implantation in patients with low anthropometry
Objective: to conduct bench tests and determine the working range of the pump speed for an implanted left ven- tricular bypass system aimed at diagnosing and treating patients with low anthropometric status.Materials and methods. The axial pump was investigated using a custom-developed hydrodynamic...
Saved in:
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | Russian |
Published: |
Federal Research Center of Transplantology and Artificial Organs named after V.I.Shumakov
2025-07-01
|
Series: | Вестник трансплантологии и искусственных органов |
Subjects: | |
Online Access: | https://journal.transpl.ru/vtio/article/view/1942 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1839573751538122752 |
---|---|
author | A. S. Buchnev A. P. Kuleshov A. A. Drobyshev V. A. Elenkin G. A. Shevchenko N. V. Grudinin |
author_facet | A. S. Buchnev A. P. Kuleshov A. A. Drobyshev V. A. Elenkin G. A. Shevchenko N. V. Grudinin |
author_sort | A. S. Buchnev |
collection | DOAJ |
description | Objective: to conduct bench tests and determine the working range of the pump speed for an implanted left ven- tricular bypass system aimed at diagnosing and treating patients with low anthropometric status.Materials and methods. The axial pump was investigated using a custom-developed hydrodynamic test bench simulating the cardiovascular system. The bench included systems for pressure and flow measurement and registration, along with software for processing both technical and biomedical parameters.Results. The operating range of the rotor speed for the STREAM CARDIO pump required to achieve a flow rate of 2.5 ± 0.5 L/min at a pressure drop of 80 ± 5 mm Hg is 8000 ± 1000 rpm, with a power consumption of 6.5 ± 1 W. |
format | Article |
id | doaj-art-97159ef7abe64d2cb8736421a3135b51 |
institution | Matheson Library |
issn | 1995-1191 |
language | Russian |
publishDate | 2025-07-01 |
publisher | Federal Research Center of Transplantology and Artificial Organs named after V.I.Shumakov |
record_format | Article |
series | Вестник трансплантологии и искусственных органов |
spelling | doaj-art-97159ef7abe64d2cb8736421a3135b512025-08-04T14:02:13ZrusFederal Research Center of Transplantology and Artificial Organs named after V.I.ShumakovВестник трансплантологии и искусственных органов1995-11912025-07-01272818810.15825/1995-1191-2025-2-81-881356Bench studies of a small axial pump for implantation in patients with low anthropometryA. S. Buchnev0A. P. Kuleshov1A. A. Drobyshev2V. A. Elenkin3G. A. Shevchenko4N. V. Grudinin5Shumakov National Medical Research Center of Transplantology and Artificial OrgansShumakov National Medical Research Center of Transplantology and Artificial OrgansShumakov National Medical Research Center of Transplantology and Artificial OrgansShumakov National Medical Research Center of Transplantology and Artificial OrgansPeoples’ Friendship University of Russia named after Patrice LumumbaShumakov National Medical Research Center of Transplantology and Artificial OrgansObjective: to conduct bench tests and determine the working range of the pump speed for an implanted left ven- tricular bypass system aimed at diagnosing and treating patients with low anthropometric status.Materials and methods. The axial pump was investigated using a custom-developed hydrodynamic test bench simulating the cardiovascular system. The bench included systems for pressure and flow measurement and registration, along with software for processing both technical and biomedical parameters.Results. The operating range of the rotor speed for the STREAM CARDIO pump required to achieve a flow rate of 2.5 ± 0.5 L/min at a pressure drop of 80 ± 5 mm Hg is 8000 ± 1000 rpm, with a power consumption of 6.5 ± 1 W.https://journal.transpl.ru/vtio/article/view/1942heart failureleft ventricular bypasshydrodynamic benchpump flow ratepump headaxial pump |
spellingShingle | A. S. Buchnev A. P. Kuleshov A. A. Drobyshev V. A. Elenkin G. A. Shevchenko N. V. Grudinin Bench studies of a small axial pump for implantation in patients with low anthropometry Вестник трансплантологии и искусственных органов heart failure left ventricular bypass hydrodynamic bench pump flow rate pump head axial pump |
title | Bench studies of a small axial pump for implantation in patients with low anthropometry |
title_full | Bench studies of a small axial pump for implantation in patients with low anthropometry |
title_fullStr | Bench studies of a small axial pump for implantation in patients with low anthropometry |
title_full_unstemmed | Bench studies of a small axial pump for implantation in patients with low anthropometry |
title_short | Bench studies of a small axial pump for implantation in patients with low anthropometry |
title_sort | bench studies of a small axial pump for implantation in patients with low anthropometry |
topic | heart failure left ventricular bypass hydrodynamic bench pump flow rate pump head axial pump |
url | https://journal.transpl.ru/vtio/article/view/1942 |
work_keys_str_mv | AT asbuchnev benchstudiesofasmallaxialpumpforimplantationinpatientswithlowanthropometry AT apkuleshov benchstudiesofasmallaxialpumpforimplantationinpatientswithlowanthropometry AT aadrobyshev benchstudiesofasmallaxialpumpforimplantationinpatientswithlowanthropometry AT vaelenkin benchstudiesofasmallaxialpumpforimplantationinpatientswithlowanthropometry AT gashevchenko benchstudiesofasmallaxialpumpforimplantationinpatientswithlowanthropometry AT nvgrudinin benchstudiesofasmallaxialpumpforimplantationinpatientswithlowanthropometry |