Golay sequences - side-lobe - canceling codes for ultrasonography

In this paper three different methods and algorithms used to calculate the pairs of Golay sequences of the different length are described. The program for simulating the pairs of the Golay sequences of different length is given. This program (using Matlab[ampersand]trade; routines) is...

Full description

Saved in:
Bibliographic Details
Main Authors: I. Trots, A. Nowicki, W. Secomski, J. Litniewski
Format: Article
Language:English
Published: Institute of Fundamental Technological Research Polish Academy of Sciences 2004-01-01
Series:Archives of Acoustics
Online Access:https://acoustics.ippt.pan.pl/index.php/aa/article/view/469
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this paper three different methods and algorithms used to calculate the pairs of Golay sequences of the different length are described. The program for simulating the pairs of the Golay sequences of different length is given. This program (using Matlab[ampersand]trade; routines) is based on the basic properties of complementary series and is simple and computationally effective. Experimental results are presented in the form, which in a clear way illustrates the resolution, depth penetration and contrast dynamics of ultrasonic images obtained using the Golay coded excitation.
ISSN:0137-5075
2300-262X