Comparative Evaluation of Classic Mechanical and Digital Goldmann Applanation Tonometers

<b>Objectives</b>: The objective of this study was to evaluate the agreement and clinical interchangeability of intraocular pressure (IOP) measurements obtained with the mechanical Haag-Streit AT900 Goldmann applanation tonometer (mGAT) and the digital Huvitz HT5000 applanation tonometer...

Full description

Saved in:
Bibliographic Details
Main Authors: Assaf Kratz, Ronit Yagev, Avner Belkin, Mordechai Goldberg, Alon Zahavi, Ivan Goldberg, Ahed Imtirat
Format: Article
Language:English
Published: MDPI AG 2025-07-01
Series:Diagnostics
Subjects:
Online Access:https://www.mdpi.com/2075-4418/15/14/1813
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:<b>Objectives</b>: The objective of this study was to evaluate the agreement and clinical interchangeability of intraocular pressure (IOP) measurements obtained with the mechanical Haag-Streit AT900 Goldmann applanation tonometer (mGAT) and the digital Huvitz HT5000 applanation tonometer (dGAT). <b>Methods</b>: This retrospective comparative study included 53 eyes of 28 patients undergoing routine ophthalmologic evaluation. Each eye underwent IOP measurement using both mGAT and dGAT in a randomized sequence. Central corneal thickness (CCT) was also recorded. Pearson’s correlation coefficient was used to determine correlation between paired IOP measurements. Bland–Altman plots were graphed for the analysis of differences for IOP between the instruments. <b>Results</b>: A total of 53 eyes of 28 patients (15 males) were included in the study. The mean age of the patients was 62.6 years. The mean mGAT and dGAT measurements were 16.3 ± 6.6 mmHg (range 9–50) and 16.4 ± 6.2 mmHg (range 8.8–45.9), respectively (<i>p</i> = 0.53). A strong, significant positive correlation was found for paired IOP measurements by the two instruments (r = 0.98; <i>p</i> < 0.0001). Bland–Altman analysis revealed 95% limits of agreement from −2.5 to +2.3 mmHg, with a small but statistically significant proportional bias favoring mGAT at higher IOP levels. Additionally, 91% of paired measurements were within ±2 mmHg. CCT-related differences were statistically and clinically insignificant. <b>Conclusions</b>: IOP measurements obtained with mGAT and dGAT were highly correlated and clinically interchangeable for the range tested. The Huvitz HT5000 may serve as a reliable alternative to the classic Goldmann tonometer in routine clinical settings.
ISSN:2075-4418