CHROMATOGRAPHIC SYSTEM SUITABILITY TESTING: INFLUENCE OF CORRECTION FACTORS AND LATE ELUTING PEAKS OF IMPURITIES ON THE REQUIREMENT FOR THE SIGNAL-TO-NOISE RATIO (REVIEW)

Introduction. In case of methods of impurities determination quite often there are problems connected both with establishment and with check of implementation of the requirement to necessary sensitivity of a chromatographic system.Text. Influence of correction factors and late eluting peaks of impur...

Full description

Saved in:
Bibliographic Details
Main Author: N. A. Epshtein
Format: Article
Language:Russian
Published: LLC Center of Pharmaceutical Analytics (LLC «CPHA») 2019-02-01
Series:Разработка и регистрация лекарственных средств
Subjects:
Online Access:https://www.pharmjournal.ru/jour/article/view/656
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Introduction. In case of methods of impurities determination quite often there are problems connected both with establishment and with check of implementation of the requirement to necessary sensitivity of a chromatographic system.Text. Influence of correction factors and late eluting peaks of impurities on the requirement for the signal-to-noise ratio (S/N) for chromatographic system suitability testing is considered.Conclusion. The recommendations are provided how to detect situations at which the requirement of S/N≥10 is insufficient for reliable detecting peaks of all impurities (related substances) and therewith how to correct concentration of test solution and/or the requirement to S/N for system suitability testing.