An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations
Microwave imaging systems show potential as replacements for commonly used stroke diagnostic systems. We developed and tested a 10-port microwave system on a liquid head phantom with ischemic and hemorrhagic strokes of varying sizes and positions. This system allows for visualization of changes in d...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2025-07-01
|
Series: | Sensors |
Subjects: | |
Online Access: | https://www.mdpi.com/1424-8220/25/14/4360 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
_version_ | 1839615226386841600 |
---|---|
author | Tomas Pokorny Jan Redr Hana Laierova Barbora Smahelova Jakub Kollar |
author_facet | Tomas Pokorny Jan Redr Hana Laierova Barbora Smahelova Jakub Kollar |
author_sort | Tomas Pokorny |
collection | DOAJ |
description | Microwave imaging systems show potential as replacements for commonly used stroke diagnostic systems. We developed and tested a 10-port microwave system on a liquid head phantom with ischemic and hemorrhagic strokes of varying sizes and positions. This system allows for visualization of changes in dielectric parameters using the TSVD Born approximation, enabling recognition of stroke position and size from the resulting images. The SVM algorithm effectively distinguishes between ischemic and hemorrhagic strokes, achieving 98% accuracy on experimental data, with 99% accuracy in ischemic scenarios and 97% in hemorrhagic scenarios. Using the TSVD Born algorithm, it was possible to precisely image changes in the absolute permittivity of different stroke locations; however, changes in stroke size were more apparent in the variations of absolute permittivity than in the reconstructed stroke size within the antenna plane. Outside this plane, changes in the S-parameters decreased depending on the distance and size of the stroke, making detection and classification more difficult. One ring of antennas around the head proved insufficient, prompting us to focus on developing a system with antennas positioned around the entire head. |
format | Article |
id | doaj-art-1bfcce4a75ce4e4fa43f122329897ba3 |
institution | Matheson Library |
issn | 1424-8220 |
language | English |
publishDate | 2025-07-01 |
publisher | MDPI AG |
record_format | Article |
series | Sensors |
spelling | doaj-art-1bfcce4a75ce4e4fa43f122329897ba32025-07-25T13:36:07ZengMDPI AGSensors1424-82202025-07-012514436010.3390/s25144360An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and LimitationsTomas Pokorny0Jan Redr1Hana Laierova2Barbora Smahelova3Jakub Kollar4Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Praha, Czech RepublicDepartment of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Praha, Czech RepublicDepartment of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Praha, Czech RepublicDepartment of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Praha, Czech RepublicDepartment of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, 160 00 Praha, Czech RepublicMicrowave imaging systems show potential as replacements for commonly used stroke diagnostic systems. We developed and tested a 10-port microwave system on a liquid head phantom with ischemic and hemorrhagic strokes of varying sizes and positions. This system allows for visualization of changes in dielectric parameters using the TSVD Born approximation, enabling recognition of stroke position and size from the resulting images. The SVM algorithm effectively distinguishes between ischemic and hemorrhagic strokes, achieving 98% accuracy on experimental data, with 99% accuracy in ischemic scenarios and 97% in hemorrhagic scenarios. Using the TSVD Born algorithm, it was possible to precisely image changes in the absolute permittivity of different stroke locations; however, changes in stroke size were more apparent in the variations of absolute permittivity than in the reconstructed stroke size within the antenna plane. Outside this plane, changes in the S-parameters decreased depending on the distance and size of the stroke, making detection and classification more difficult. One ring of antennas around the head proved insufficient, prompting us to focus on developing a system with antennas positioned around the entire head.https://www.mdpi.com/1424-8220/25/14/4360brain strokemicrowave devicesTSVD Born approximationSVM |
spellingShingle | Tomas Pokorny Jan Redr Hana Laierova Barbora Smahelova Jakub Kollar An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations Sensors brain stroke microwave devices TSVD Born approximation SVM |
title | An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations |
title_full | An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations |
title_fullStr | An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations |
title_full_unstemmed | An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations |
title_short | An Experimental 10-Port Microwave System for Brain Stroke Diagnosis—Potentials and Limitations |
title_sort | experimental 10 port microwave system for brain stroke diagnosis potentials and limitations |
topic | brain stroke microwave devices TSVD Born approximation SVM |
url | https://www.mdpi.com/1424-8220/25/14/4360 |
work_keys_str_mv | AT tomaspokorny anexperimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT janredr anexperimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT hanalaierova anexperimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT barborasmahelova anexperimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT jakubkollar anexperimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT tomaspokorny experimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT janredr experimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT hanalaierova experimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT barborasmahelova experimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations AT jakubkollar experimental10portmicrowavesystemforbrainstrokediagnosispotentialsandlimitations |